High Precision Calibration Kit, 67GHz, 1.85mm

SCKCL67-1.85 mechanical calibration kit contains precision standard devices to calibrate network analyzers in the 1.85mm interface. This kit also contains adapters to change the sex of the test port, and a torque wrench to ensure a repeatable reliable connection.

This calibration kit can use the calibration data of 85058B.

Key Features

DC to 67GHz frequency coverage

Adapters and torque wrench included

> 500 times cycles

Calibration standards to perform full two-port calibration

Application

Site Master

Cable and Antenna Analyzer

Vector Network Analyzer (VNA)

Unit Price:

SCKCL67A1-1.85: $11,740 (1-5pcs)

SCKCL67A2-1.85: $9,240 (1-5pcs)

NOTICE: Online purchases are temporarily closed. Sorry for the inconvenience.

Detailed

SCKCL67A1-1.85

Model Frequency Specifications List Qty.
SCKCL67A1-1.85

 

(Calibration Level)

DC – 67GHz Open: ± 2.0°

(10MHz-50GHz)

Open: ± 3.0°

(50MHz-67GHz)

 

Short1:±2°
(10MHz-67GHz

Short2:±2.5°
(35GHz-67GHz

Short3:±2.5°
(45GHz-67GHz

Short4:±2.5°

(35GHz-67GHz)

 

Load: VSWR

≥36dB
(10MHz-10GHz)
≥35dB
(10GHz-20GHz)
≥31dB
(20GHz-35GHz)

 

Adaptor: VSWR ≥33dB
(10MHz-4GHz)
≥24dB
(4GHz-26.5GHz)
≥22dB
(26.5GHz-50GHz)
≥20dB
(50GHz-67GHz)

Open: 1.85mm (M) 1
Open: 1.85mm (F) 1
Short: 1.85mm (M) 4
Short: 1.85mm (F) 4
Load: 1.85mm (M) 1
Load: 1.85mm (F) 1
1.85mm(f) – 1.85mm(f) adaptor 1
1.85mm(m) – 1.85mm(m) adaptor 1
1.85mm(m) – 1.85mm(f) adaptor 1
Torque wrench 1

SCKCL67A2-1.85

Model Frequency Specifications List Qty.
SCKCL67A2-1.85

 

(Calibration Level)

DC – 67GHz Open: ± 3.0°

 

Short: ± 2.5°

 

Load: VSWR

≥30dB
(DC-26.5GHz)
≥28dB
(26.5GHz-50GHz)
≥25dB
(50GHz-67GHz)

 

Adaptor: VSWR ≥33dB
(DC-4GHz) ≥24dB
(4GHz-26.5GHz)
≥22dB
(26.5GHz-50GHz)
≥20dB
(50GHz-67GHz)

Open: 1.85mm (M) 1
Open: 1.85mm (F) 1
Short: 1.85mm (M) 1
Short: 1.85mm (F) 1
Load: 1.85mm (M) 1
Load: 1.85mm (F) 1
1.85mm(f) – 1.85mm(f) adaptor 1
1.85mm(m) – 1.85mm(m) adaptor 1
1.85mm(m) – 1.85mm(f) adaptor 1
Torque wrench 1

Unboxing Video

How to Place an Order ?

Common Q&A

  • Q: Can I use it in Keysight VNA?

    A: Yes, you can, insert the U Disk which we provided to Keysight VNA, import the database, then you can use it successfully.

  • Q: Can I use it to R&S, Anritsu or other brand VNA?

    A: Yes, you can. We will provide you a database, you just need to input the data to your VNA at the first time, then can use it freely in future.

  • Q: Can I use it in my Sitemaster or Cable and Antenna Testers?

    A: Yes, you can. We will provide you a database, you just need to input the data to your VNA at the first time, then can use it freely in future.

  • Q: What is the warranty time?

    A: We guarantee at least 500 times, In case there is quality issue, please contact with us for free repair within one year.

  • Q: Who will afford shipping cost for return?

    A: Shipping from your side to our factory need to pay by yourself, we will pay shipping cost when shipping back.

  • Q: Return Policy

    A: If there is no quality problem, we do not accept returns and refunds.

  • Q: Refund Policy

    A: If the return is caused by the product quality or the mistake from us, you will guaranteed to get the refund by paypal.